Sample Pipetting Unit
Sample Rack :80 sample positions(tubes 12mm)
Primary Sample Tubes :12*75mm
Pipetting:10ul~1000ul, (100ul) CV<0.5%
Dispensing System
1 aspirating-dispensing probe with X-Y-Z movement; dilution system with a syringe of 1000ul, continuous flow washing station
.
Reagent Pipetting Unit
Reagent Racks :8 special reagent racks,1 general reagent racks
Removable Trays :Modular racks to hold calibrators, controls and reagents for 8 test on line
Pipetting :10ul~1000ul,1ul stepping
Pipetting precision :(100ul) CV<0.5%
Time to Dispense:4 minutes adding 96 reagents
Incubation Unit
Incubators :2 independent units
Temperature Range:RT to 45 °C
Temperature Accuracy:±0.5°C
Temperature Fluctuation:±0.5°C
Washing Unit
Programmable : Yes(options & parameters)
Wash Containers : 2 wash buffers at 2.0L, with level-sensing
Washing System:Two 8 nozzle manifolds, one to dispense and the other one to aspirate
Washing Residual : <2μl
Reading Unit
Reading System : 8 independent photometric channels with mono and dichromatic reading capability
Photometric Range : 0.000 ~3.000 OD
Spectral Range : 400-700nm
Optical Filters : 2 precision filters (450,630nm); 2 more on request (4 filters max, 400-700nm)
Software Unit
Software System : Windows 7/8/10 or above, LIS system available
Function Application
Work list set up, patient reports and result filing; calibration curve storage for each carried out test
Data Reduction
cut-off (qualitative)
Physical Specification
Package Size(W*D*H) : 920*700*1360mm
Gross Weight : 110kg
Power Supply :AC220V+10%,50/60Hz;110V+10%,60Hz; 500W