Tel: +86-020-39160381/39154381
Your position: Home > Laboratory equipment

MY-B034D Specific protein analyzer

MY-B034D Specific protein analyzer
Item Code: MY-B034D
    Please select the information you wantX
Shipping Cost:
to
Estimated Delivery Time:
Please select the country you want to ship from
+
-
Set/Sets 100 in Stock
  • Description
Specifications
Parameters
1)The analyzer geometry: length × width × height = 250 mm × 190 mm × 85 mm.
2) The analyzer weighs about: 2kg.
3) analyzer using a laser diode as a light source, the operating wavelength of 650nm.
4) turbidity measurement range 0 ~ 300FTU.
5) Injection mode: disposable semi-micro cuvette capacity of 2.5ml, PMMA materials, optical path 10mm.
6) Display: 16 × 2 character dot matrix LCD display, can display English numbers.
7) Print: Built 57mm thermal printer can print measurement results.
8) Item Storage: parameter test items stored in the IC card, the reading measured by the built-in card reader.
9) Keyboard: 5 keys touch keyboard.
10) energy consumption: Analyzers maximum input power does not exceed 40VA.
Third, the requirement
1, the working conditions:
a) Power supply voltage: A.C.220V ± 22V, 50Hz ± 1Hz;
b) Input power: ≤30VA;
c) Preheat: ≥10min;
d) Ambient temperature: 15 ℃ ~ 35 ℃;
e) Relative humidity: ≤95%;
f) Atmospheric pressure: 700hPa ~ 1060hPa;
No nearby g) the instrument of strong magnetic interference, no vibrations, no direct sunlight.
2, the main indicators:
a) Stray light: no turbidity measuring water turbidity should ≤1 (FTU);
b) scattered light intensity Repeatability: scattering intensity repeatability CV≤3%;
c) temperature accuracy and temperature fluctuations: The analyzer cuvette room thermostat temperature 37 ℃, in fact, the temperature accuracy temperature measurement error shall be less than ± 0.5 ℃, short-term fluctuations in the measured temperature is less than ± 0.2 ℃;